Stationary gas analyzers IGM-13M are designed for automatic, continuous measurement of the concentration of explosive gases, toxic gases, oxygen and carbon dioxide in the surrounding atmosphere.
Hazardous areas according to the explosion protection marking 1ExdiaIIC T6 X, in which the formation of explosive mixtures of gases and vapors with air is possible.
The devices can be used in the coal, oil, gas and energy industries.
IGM-13M is made in a metal case with a transparent window for indication elements, or without a window, but with a light guide for indication. The housing contains cable glands, a cover for access to external commutation terminals, and a metal optical sensor module.
Sampling method | diffusion |
---|---|
Operating principle | optical, electrochemical, photoionization. |
Determined components: | combustible gases, CO2 (optical sensor); toxic gases, H2, 02 (electrochemical sensor); MPC of harmful and hydrocarbon gases (photoionization sensor). |
Operating conditions: - ambient air temperature, º С - relative humidity (without condensation),% - atmospheric pressure, kPa | from - 60 to +60 from 15 to 95 from 80 to 120 |
Overall dimensions, mm, no more | 225х125х95 |
Weight, kg, no more | 1,5 Al / 3,5 AISI316 |
Body material | aluminum stainless steel (optional) |
Output signals | RS-485 protocol MODBUS RTU, 4-20 mA, 3 relays, HART |
Average service life, years, not less | ten |
Sheath protection degree | IP67 |
Climatic modification according to GOST 15150 | O1 |
Explosion protection marking | 1Exd ia IIC T6, 10 |
Calibration interval, years | 2 (optical sensors) 1 (electrochemical and photoionization sensors). |
Power Options | |
DC voltage, V | from 12 to 32 |
Gas analyzer consumed, no more, W | 0.6 |