FLOW ANALYZER AQUATOS-UV-254
Designed for continuous determination of chemical oxygen demand, biological oxygen demand, total organic carbon in water.
The measuring principle is based on the absorption of UV radiation by molecules of aromatic hydrocarbons at a wavelength of 254 nm.
Turbidity, suspended solids or impurities in the flow cell are automatically compensated for by differential measurement with a second sensor with a different wavelength (590 nm).
ADVANTAGES COMPARED TO SUBMERSIBLE SENSORS:
Possibility of two-stream analysis;
Automatic calibration and verification of the correctness of the readings, which is not possible with submersible sensors;
Automatic cleaning with a programmable cycle;
Wide measurement range using the built-in dilution module.
Method: | Measurement of UV absorbance at 254 nm |
Measurement interval: | Programmable, minimum 60 sec |
Error: | ± 2% of absorbance for samples with turbidity < 100 FUU |
Nutrition: | 110-220 VAC, 50-60 Hz, 80 Volt-Amp. |
Mounting: | Wall or pole mounting |
Ambient temperature: | 5 - 45 ° C |
Measuring cell diameter: | 16 mm (optical path length 12 mm) |
Protection class: | IР54 |
Housing: | Cold rolled epoxy powder coated steel |
Dimensions: | 380x600x210 mm (LxHxW) |
Weight: | Approx. 20 kg |
Access settings: | Two levels of password access |
Output signals: | 2х4-20 mA, ModBUS RTU RS485 |
4 programmable relays (signal on failure, on , no flow 1, no flow 2, exceeding the preset concentration value, etc.) | |
Data storage: | Built-in recorder with USB port |
Automatic functions: | Calibration, validation, cleaning |
Sample requirements: | |
Temperature: | 5-45 ° C |
Sample inlet pressure: | Atmospheric |
Sample outlet pressure: | Atmospheric (drain to drain) |
Sample feed rate into fast loop reservoir: | 100-500 ml / min |
Connecting dimensions: | Sample to the tank inlet with quick loop - flexible tubing 6 mm OD, drain tubing 12 mm OD |